Gesamtkatalog

Laden Sie sich unseren umfangreichen Katalog herunter und entdecken Sie auch viele weitere Produkte von REO.

Isolating transformer REOMED II

Transformers for medical technology

DESCRIPTION

The REOMED II isolation transformer is a reliable leakage current limiter in medical systems.
Voltage-resistant materials – combined with high-quality processing – provide safe isolation from the mains input side. The inbuilt isolation transformers are designed for low internal energy loss, resulting in very low no-load losses. With regards to the input power of REOMED II, values of <1 % have been reached.
A circuit breaker ensures that the transformers are protected against overload and short-circuiting on both the input and output sides.
The mains input side must be operated with a mains switch with a green light. A two-pole circuit breaker on the mains side and a one-pole circuit breaker on the output side provide protection.
There is no need to keep replacement fuses available, and there is also no risk of confusion as a result.
Furthermore, a temperature switch is integrated into the power isolation transformer, providing additional safety if there is no internal cooling system.

ADVANTAGES

  • Very good efficiency
  • low-noise
  • Very low magnetic stray field
  • Easy connectivity
  • Compact dimensions
  • Sturdy aluminium housing
  • Mains switch with green light
  • Low overall weight
  • Integrated circuit breakers
  • IEC 60601-1:2005/ AMD1:2012; ANSI/ AAMIES
    60601-1:2005/ CR/ 2012 CAN/ CSA C22.2 No.
    60601-1:14 (medical technology) USA and Canada IEC
    60601-1-2:2014 (Partly); EN/ IEC 61000-3-2:2014;
    EN/ IEC 61000-3-3:2013

 

Input voltage 230 V
Output voltage 230 V
Nominal power consumption 660 – 2000 VA
Enclosure protection class IP 20
Weight 7.7 – 18.0 kg
Primary circuit breaker 4 – 12 A
Secondary circuit breaker 3 – 10 A
Earth leakage current at 254V / 50/60Hz < 500µA
Number of output sockets 6x according to IEC 320
Test voltage 4 kVac (between primary and secondary winding)
Max.
Ambient temperature
40 °C
Insulation resistance > 2 MΩ
Protective conductor resistance < 0,1 Ω

All devices have an inrush current limiter (NTC or electronic), potential equalization in accordance with DIN 42801, a primary mains cable and a protective temperature limiter. The devices can be wall, table or floor mounted