Transformers for medical technology
DESCRIPTION
The REOMED II isolation transformer is a reliable leakage current limiter in medical systems.
Voltage-resistant materials – combined with high-quality processing – provide safe isolation from the mains input side. The inbuilt isolation transformers are designed for low internal energy loss, resulting in very low no-load losses. With regards to the input power of REOMED II, values of <1 % have been reached.
A circuit breaker ensures that the transformers are protected against overload and short-circuiting on both the input and output sides.
The mains input side must be operated with a mains switch with a green light. A two-pole circuit breaker on the mains side and a one-pole circuit breaker on the output side provide protection.
There is no need to keep replacement fuses available, and there is also no risk of confusion as a result.
Furthermore, a temperature switch is integrated into the power isolation transformer, providing additional safety if there is no internal cooling system.
Voltage-resistant materials – combined with high-quality processing – provide safe isolation from the mains input side. The inbuilt isolation transformers are designed for low internal energy loss, resulting in very low no-load losses. With regards to the input power of REOMED II, values of <1 % have been reached.
A circuit breaker ensures that the transformers are protected against overload and short-circuiting on both the input and output sides.
The mains input side must be operated with a mains switch with a green light. A two-pole circuit breaker on the mains side and a one-pole circuit breaker on the output side provide protection.
There is no need to keep replacement fuses available, and there is also no risk of confusion as a result.
Furthermore, a temperature switch is integrated into the power isolation transformer, providing additional safety if there is no internal cooling system.
ADVANTAGES
- Very good efficiency
- low-noise
- Very low magnetic stray field
- Easy connectivity
- Compact dimensions
- Sturdy aluminium housing
- Mains switch with green light
- Low overall weight
- Integrated circuit breakers
- IEC 60601-1:2005/ AMD1:2012; ANSI/ AAMIES
60601-1:2005/ CR/ 2012 CAN/ CSA C22.2 No.
60601-1:14 (medical technology) USA and Canada IEC
60601-1-2:2014 (Partly); EN/ IEC 61000-3-2:2014;
EN/ IEC 61000-3-3:2013
Input voltage | 230 V |
Output voltage | 230 V |
Nominal power consumption | 660 – 2000 VA |
Enclosure protection class | IP 20 |
Weight | 7.7 – 18.0 kg |
Primary circuit breaker | 4 – 12 A |
Secondary circuit breaker | 3 – 10 A |
Earth leakage current at 254V / 50/60Hz | < 500µA |
Number of output sockets | 6x according to IEC 320 |
Test voltage | 4 kVac (between primary and secondary winding) |
Max. Ambient temperature |
40 °C |
Insulation resistance | > 2 MΩ |
Protective conductor resistance | < 0,1 Ω |
All devices have an inrush current limiter (NTC or electronic), potential equalization in accordance with DIN 42801, a primary mains cable and a protective temperature limiter. The devices can be wall, table or floor mounted